Draeger Infinity Acute Care System Monitoring Solution

Company
Draeger Medical Systems, Inc.
Recall Initiated
December 10, 2013
Posted
January 10, 2014
Terminated
May 20, 2015
Recall Number
Z-0714-2014
Quantity
439
Firm Location
Telford, PA

Reason for Recall

After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Distribution

US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.

Lot / Code Info

with software versions VG2.0.3 and higher.

Root Cause

Software design

Action Taken

An Urgent Medical Device Recall Letter, dated December 2013, was sent to Users. Once available Drager will update components of the IACS Monitoring Solution System.