Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Company
Draeger Medical Systems, Inc.
Recall Initiated
March 5, 2014
Posted
April 10, 2014
Terminated
October 31, 2014
Recall Number
Z-1440-2014
Quantity
3
Firm Location
Telford, PA

Reason for Recall

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Distribution

USA Nationwide Distribution in the states of WI and NY.

Lot / Code Info

Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003,  ASEK-0004, ASEK-0005

Root Cause

Process control

Action Taken

Drager sent an Urgent Medical Device Recall letter dated February 2014 to affected customers. The letter identified the affected device, problem and actions to be taken. The notification directed customers to perform manual ventilation in the event of a failure. Drager plans to repair the affected devices free of charge. If you have any questions regarding the Recall letter please call 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of your Drger ventilator please contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt).