Aesculap Columbus Revision CoCr 15mm distal and posterior spacers, sizes F5-F7
- Company
- Aesculap Inc
- Recall Initiated
- September 19, 2013
- Posted
- December 13, 2013
- Terminated
- December 22, 2014
- Recall Number
- Z-0492-2014
- Quantity
- 24
- Firm Location
- Center Valley, PA
Reason for Recall
The Columbus Revision CoCr 15 mm distal and posterior implant spacers, sizes F5-F7, were not compatible.
Distribution
Distributed in FL and MD.
Lot / Code Info
lots: 51468773, 51447588, 51468302, 51503760, 51571394, 51585136, 51447597
Root Cause
Component change control
Action Taken
Domestic consignees were contacted by phone and a recall notification letters was also sent to these Users on/about September 26, 2013.
More recalls by Aesculap Inc
Brand Name: AESCULAP
Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM
Model/Catalog Number: FF3...
Mar 2, 2026
Brand Name: Aesculap
Product Name: SEALING CAP FOR 10/12MM TROCARS
Model/Catalog Number: EK085P
Soft...
Oct 24, 2025
Brand Name: Aesculap
Product Name: SEALING UNIT F/10/12MM TROCARS W.REDUCER
Model/Catalog Number: EK...
Oct 24, 2025