RecallDepth

Brand Name: AESCULAP Product Name: SEALING UNIT FOR 10/12MM TROCARS Model/Catalog Number: EK086P Software Version: N/A Product Description: SEALING UNIT FOR 10/12MM TROCARS Component: N/A

Company
Aesculap Inc
Recall Initiated
September 24, 2025
Posted
October 24, 2025
Recall Number
Z-0311-2026
Quantity
26 units
Firm Location
Center Valley, PA

Reason for Recall

It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Distribution

Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.

Lot / Code Info

Model No EK086P; UDI-DI: 04046963425272; Lots 52968398, 52975386, 52988022, 52993532, 53008999, 53009000, 53009001, 53009417

Root Cause

Under Investigation by firm

Action Taken

On September 24, 2025 URGENT MEDICAL DEVICE RECALL NOTIFICATION letters were sent to customers. Actions to be taken. 1. Review the recall notice and ensure all users in the organization are notified. Forward this notification to customers. Remove affected units from inventory, quarantine the product but do not destroy it. Complete and return the included response form. Return affected units to the firm. Credit of the product will be issued. An alternative disposable universal seal is available and can be ordered.