Vertier Surgical Table, an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system. Part Numbers: 0788100000 Standard table, 0788100001 Standard table with 5th wheel, 0788100000S SIDNE-enabled standard table, and 0788100001S SIDNE-enabled 5th wheel table

This recall has been terminated (originally issued August 26, 2009).

Company
Stryker Communications Corp
Recall Initiated
November 13, 2008
Posted
August 26, 2009
Terminated
August 28, 2009
Recall Number
Z-1899-2009
Quantity
31
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Surgical tables were produced using an unapproved tilt frame bushing.

Distribution

Worldwide Distribution -- United States (OR, IL, WA, CA, AR, TX, UT, SC and GA) and Canada.

Lot / Code Info

Lot Numbers: 080111-43525, 080111-43526, 080111-43527, 080111-43528, 080118-43529, 080118-43530, 080118-43531, 080118-43532, 080118-43533, 080118-43537, 080118-43538, 080811-54955, 080811-54956, 080811-54957, 080811-54958, 080811-54959, 080811-54961, 080811-54962, 080811-54963, 080811-54964, 080811-54965, 080811-54966, 080811-54967, 080811-54968, 080811-54985, 080811-54986, 080811-54987, 080811-54988, 080811-54989, 080811-54998 and 080811-55002.

Root Cause

Nonconforming Material/Component

Action Taken

Stryker Communications issued an "Urgent: Medical Device Recall" notice dated November 13, 2008 informing Consignees of the affected product. The firm intends to contact each facility to coordinate service by a trained representative. For further information, contact Stryker at 1-972-410-7100.

More recalls by Stryker Communications Corp