Measles IgG Enzyme Immunoassay Test Kit, DIAMEDIX. The assay is intended for the semi-quantitation of human IgG antibodies to measles virus in human serum by indirect immunoassay to aid in the assessment of the patient's immunological response to measles virus, to determine the immune status of individuals and, when evaluating paired sera, as an aid in the diagnosis of measles infection.

This recall has been terminated (originally issued January 7, 2009).

Company
Diamedix Corporation
Recall Initiated
October 10, 2008
Posted
January 7, 2009
Terminated
May 12, 2010
Recall Number
Z-0607-2009
Quantity
974Test Kits
Firm Location
Miami, FL
Official Source
View on FDA website ↗

Reason for Recall

Possible false-negative test results: The Sample A Diluent Lot # 50768 assigned to Is-Measles IgG Lot #'s 41108 and 41108B has higher conductivity than normal, which can sporadically interfere with the MAGO Plus/Aptus sensing of and picking up of samples. As a result, if the sample is not picked up/processed, it could lead to the possibility of obtaining false negative results in some cases.

Distribution

Worldwide Distribution --- including USA and country of England.

Lot / Code Info

Lot number 41108 and 41108B

Root Cause

Other

Action Taken

Notification of all domestic customers via Certified Mail. Notification of the one international customer via Federal Express.