The Is-Legionella IgG/IgM/IgA test kits are packaged in cardboard boxes with individual components secured within a foam insert.

This recall has been terminated (originally issued February 15, 2005).

Company
Diamedix Corporation
Recall Initiated
December 23, 2004
Posted
February 15, 2005
Terminated
May 3, 2012
Recall Number
Z-0541-05
Quantity
178
Firm Location
Miami, FL
Official Source
View on FDA website ↗

Reason for Recall

The kits belonging to lot #s 30804/30804Z and 91803Z may lose activity, after one or two usages, causing the OD values for the Cut-Off Calibrator and/or Positive Control to be recovered below their specified ranges.

Distribution

Distributed to Domestic customers only.

Lot / Code Info

Lot # 91803Z and 30804/30804Z

Root Cause

Other

Action Taken

Direct customers were notified by letter on 12/23/2004 by Federal Express.100% of the total customers will be contacted. The letter instructed the consignees to discontinue use immediately and arrange to return the product to Diamedix.