RecallDepth

Kimberly-Clark Microcuff Endotracheal Tube, Pediatric Oral/Nasal Magill, sizes 3.0mm, 3.5mm, 4.0mm, and 4.5mm, Rx Only, Sterile, Manufacturer: Kimberly-Clark NV Belgicastraat 13, 1930 Zaventem, Belgium, Made in Malaysia

This recall has been terminated (originally issued August 8, 2007).

Company
Kimberly-Clark Corporation
Recall Initiated
June 4, 2007
Posted
August 8, 2007
Terminated
June 12, 2008
Recall Number
Z-1143-2007
Quantity
4376 cases (43760 units/total for both products)
Firm Location
Roswell, GA
Official Source
View on FDA website ↗

Reason for Recall

The tubes may kink during use which may impede or prevent ventilation.

Distribution

Worldwide, including USA, Canada, Europe, Australia, Chile, Hong Kong, Singapore, Thailand, India, Malaysia and Saudi Arabia.

Lot / Code Info

Product code 35111-07, 35112-07, 35113-07 and 35114-07, All Lots.

Root Cause

Other

Action Taken

Consignees were notified by letter sent via overnight mail on/about June 4, 2007. A follow-up letter was sent to consignees via fax on June 5, 2007. Consignees were instructed to cease distributing the product and return all affected product in inventory to Kimberly-Clark. Distributors were instructed to notify their accounts of the recall. A Product Recall Response Sheet (Packing Slip) was attached to the recall notice.