Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system, indicated for monitoring and controlling patient temperature.

This recall has been terminated (originally issued November 29, 2012).

Company
Kimberly-Clark Corporation
Recall Initiated
September 16, 2010
Posted
November 29, 2012
Terminated
December 20, 2012
Recall Number
Z-0451-2013
Quantity
400 units
Firm Location
Roswell, GA
Official Source
View on FDA website ↗

Reason for Recall

Error 85 alarm on the Model 1000 Control Unit was occasionally triggered inappropriately.

Distribution

Nationwide Distribution including AZ, AK, CA, CO, DE, DC, FL, GA. ID, IL, IN, IA, KS, KY, LA, MD, ME MA, MS, MO, MT, NE, NJ, NM, NV, NY, NC, OH, OR, PA, RI, SC, TN, TX, VA, WV, and WI

Lot / Code Info

Serial Number - SA09043017, SA09043019, SA0904320, SA09043018, SA09043015, SA09043012, SA09043016, SA09043014, SA09043013, SA09043009, SA09043011, SA09043010, SA09043008, SA09043007, SA09043006, SA09043003, SA09043005, SA09043004, SA09043002, SA09043001, SA09020013, SA09020012, SA09020014, SA09020011, SA09020010, SA09020009, SA0902008, SA09020006, SA09020007, SA08343100, SA09020004, SA09020005, SA09020002, SA09020003, SA09020001, SA08343099, SA08343098, SA08343097, SA08343096, SA08343095, SA08343094, SA08343093, SA08343092, SA08343091, SA08343090, SA08343089, SA08343087, SA08343088, SA08343086, SA08343085, SA08343083, SA08343084, SA08343053, SA08343073, SA08288014, SA08288002, SA08343054, SA08343082, SA08343081, SA08343080, SA08343079, SA08343078, SA08343077, SA08165007, SA08343076, SA08343075, SA08343074, SA08343072, SA08343055, SA08288007, SA08343071, SA08343070, SA08343069, SA08343068, SA08343067, SA08343066, SA08343065, SA08343064, SA08343063, SA08343059, SA08343058, SA08343062, SA08343061, SA08343060, SA08343056, SA08343057, SA08288012, SA08288017, SA08288016, SA08288015, SA08288013, SA08288011, SA08288009, SA08288006, SA08288003, SA08288010, SA08288008, SA08288004, SA08288005, SA08288001, SA08165013, SA08165014, SA08165015, SA08165018, SA08165016, SA08165017, SA08165006, SA08165005, SA08165011, SA08165012, SA08165010, SA08165009, SA08165004, SA08165008, SA08165002, SA08165003, SA08144035, SA08144030, SA08165001, SA08144034, SA08144033, SA08144032, SA08144031, SA08144029, SA08144028, SA08144025, SA08144027, SA08144021, SA08144026, SA08144024, SA08144023, SA08144022, SA08136010, SA08136006, SA08136003, SA08136009, SA08136005, SA08136004, SA08136007, SA08136001, SA08136002, SA08136008, SA08094129, SA08094137, SA08094141, SA08094136, SA08094150, SA08094153, SA08094152, SA08094151, SA08094149, SA08094145, SA08094148, SA08094147, SA08094143, SA08094146, SA08094142, SA08094144, SA08094140, SA08094139, SA08094138, SA08094134, SA08094130, SA08094132, SA08094135, SA08094131, SA08094133, SA08094127, SA08094128, SA08094125, SA08094126, SA08094124, SA08017044, SA08017049, SA08017045, SA08017040, SA08017034, SA08017035, SA08017041, SA08017047, SA08017042, SA08017043, SA08017048, SA08017039, SA08017046, SA08017038, SA08017037, SA08017030, SA08017036, SA08017028, SA08017029, SA08017027, SA08017033, SA08017031, SA08017032, SA07344022, SA07344029, SA07344028, SA07344026, SA07344027, SA07344030, SA07344021, SA07344023, SA07344024, SA07344025, SA07320002, SA07320001, SA07320014, SA07320015, SA07320013, SA07320011, SA07320012, SA07320006, SA07320005, SA07320008, SA07320010, SA07320009, SA07320007, SA07320004, SA07320003, SA07261006, SA07257008, SA07261004, SA07261007, SA07261002, SA07257004, SA07257007, SA07257003, SA07257002, SA07261005, SA07261008, SA07261003, SA07261001, SA07257005, SA07257006, SA07257001, SA07250005, SA07250004, SA07250002, SA07250003, SA07250001, SA07228007, SA07228008, SA07228001, SA07228002, SA07228003, SA07228004, SA07228005, SA07228006, SA07180006, SA07180005, SA07180004, SA07180003, SA07180002, SA07180001, SC07171006, SC07171005, SC07171004, SC07171003, SC07171002, SC07171001

Root Cause

Component design/selection

Action Taken

Kimberly Clark sent an Urgent Field Corrective Action letter dated September 16, 2010, to all affected customers. The letter identified the product, the problem, and the aciton to be taken by the customer. Customers were asked to confirm whether their facility as the affected product by faxing the completed Field Corrective Action Response Form to 920-380-6682. Kimberly-Clark will arrange to perform the upgrade at the customer's facility. Customers with questions should call the Field Corrective Action Coordinator at 770-587-7400. For questions regarding this recall call 770-587-8393.