Nemio XG, ultrasound system, SSA-590A with version 1.5 software, Toshiba American Medical Systems, Inc., Tusin, CA.

Company
Toshiba American Med Sys Inc
Recall Initiated
December 19, 2006
Posted
May 5, 2007
Terminated
February 15, 2012
Recall Number
Z-0771-2007
Quantity
37
Firm Location
Tustin, CA

Reason for Recall

Transducer Surface Temperature : It was found during testing that when using the system in the 2DCDI + PW mode, with the Scale Graph function active, that the transducer surface temperature could exceed the 27 degrees Celsius limit as specified in the IEC safety standard.

Distribution

Nationwide Distribution including the states of CA, FL, GA, ID, MA, MN, NC, NY, PA, RI, TN, TX, WA, and WI.

Lot / Code Info

Serial #: A1A0642100; A1A0642102; A1B0662183; A1B0662184; A1B0662256; A1B0662257; A1B0662258; A1B0662327; A1B0662328; A1B0662330; A1B0662331; A3A0652115; A3B0662284; A3B0662285; A3B0662286; A3B0662287; A3B0662288; A3B0662289; A3B0662290; A3B0662291; A3B0662292; A3B0662293; A3B0662371; A3B0662372; A3B0662373; A3B0662374; A3B0662375; A3B0662376; A3B0662377; A3B0662378; A3B0662379; A3B0662380; A3B0672415; A3B0672416; A3B0672417; A3B0672418; A3B0672419;

Root Cause

Other

Action Taken

Notification letters were sent to customers on December 19, 2206. Customers were instructed to avoid using Scale Graph function with live images, and instructed to call up the Scale Graph after images are frozen. Customers were informed that a Toshiba Customer Service Engineer will contact them to schedule a software upgrade when software becomes available. If customers have any questions they were instructed to contact their Toshiba Customer Service Engineer or call (800) 421-1968 and ask for the Regulatory Affairs Department. Customers were also instructed to distribute this information to all users of the product and also with the Bio-medical Department at their facility.