Toshiba America Medical Systems (TAMS) Aquilion 64 CT system, computed tomography x-ray system.

Company
Toshiba American Med Sys Inc
Recall Initiated
October 1, 2007
Posted
September 29, 2008
Terminated
July 16, 2012
Recall Number
Z-1708-2008
Quantity
1203 total
Firm Location
Tustin, CA

Reason for Recall

Excessive radiation: Intermittent shifting of CT numbers causes the actual dose to the patient to be greater than expected, when Sure Exposure option is used.

Distribution

Worldwide distribution, including USA and Japan.

Lot / Code Info

All codes

Root Cause

Device Design

Action Taken

Toshiba sent a letter to customers in July 2007, and sent field service representatives to update the Data Aquisition System in September/October 2007.