NICO Model 7300 Cardiopulmonary Management System

This recall has been terminated (originally issued September 29, 2006).

Company
Respironics Novametrix, Llc.
Recall Initiated
August 11, 2006
Posted
September 29, 2006
Terminated
June 12, 2012
Recall Number
Z-1535-06
Quantity
2,432 units
Firm Location
Wallingford, CT
Official Source
View on FDA website ↗

Reason for Recall

Audible alarm may not sound during an alert condition

Distribution

Worldwide distribution --- USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Bulgaria, Bahrain, Brazil, Canada, Canton of Aargau Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Algeria, Ecuador, Egypt, Spain, Finland, France, Great Britain, Guam, Hong Kong, Hungary, Ireland, India, Italy, Jamaica, Jordon, Japan, Korea, Kuwait, Lebanon, Lithuania, Morocco, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Peru, Phillippines, Puerto Rico, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovania, Slovakia, Thailand, Turkey, Taiwan, Uruguay, Venezuala, and South Africa.

Lot / Code Info

All serial numbers beginning with 138

Root Cause

Other

Action Taken

Respironics notified consignees of the problem with the audible alarm by certified letter on 8/28/06. The correction requires replacement of a speaker upgrade kit.