RecallDepth

FloTrak Elite modules FloTrak Elite P/N 1044971 FloTrak Elite P/N 1051005

This recall has been terminated (originally issued September 30, 2011).

Company
Respironics Novametrix, Llc.
Recall Initiated
July 29, 2011
Posted
September 30, 2011
Terminated
May 23, 2012
Recall Number
Z-3288-2011
Quantity
41 units
Firm Location
Wallingford, CT
Official Source
View on FDA website ↗

Reason for Recall

Pressure lines maybe occluded and display incorrect tidal volumes.

Distribution

Worldwide Distribution -- USA, Belgium, Canada, Chile, Columbia, Egypt, India, Italy, Spain, Saudi Arabia Thailand, GB, and Russia.

Lot / Code Info

Serial Numbers 525 539 146 155 171 173 178 611 612 150 165 231 234 175 235 236 237 238 239 240 356 376 389 414 000475 483 000496 502 510 511 686 693 700 723 724 725 167-000385 167-000487 232 000372, and 000834.

Root Cause

Manufacturing material removal

Action Taken

Respironics Novametrix, LLC on July 29, 2011 notified accounts via e-mail or tracked receipt, an Urgent - Medical Device Recall letter to notify customers of the issue and to instruct them on conditions for continued use. Monitors will need to be returned for functional performance testing at their facility. Customers can contact Respironics US Customer Service at 1-877-387-3311 for customers residing in the USA or +011 (203)697-6348 for International customers, if they have any questions or need additional information.