Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT110
This recall has been terminated (originally issued March 29, 2006).
- Company
- Fisher & Paykel Healthcare Inc
- Recall Initiated
- December 20, 2005
- Posted
- March 29, 2006
- Terminated
- April 19, 2012
- Recall Number
- Z-0694-06
- Quantity
- 2699 boxes (26,990 pcs)
- Firm Location
- Laguna Hills, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.
Distribution
Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution
Lot / Code Info
Affected lots for all catalogue numbers: 041101-050331(inclusive).
Root Cause
Other
Action Taken
On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.
More recalls by Fisher & Paykel Healthcare Inc
Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT211
Mar 29, 2006
Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT210
Mar 29, 2006
Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT111
Mar 29, 2006
RT130 and RT131 Neonatal/ Infant Breathing Circuits, heated wire. For use with MR850 Respiratory Hum...
Feb 18, 2005
Phototherapy Lamp, Model 9001W501JE
Dec 18, 2003