Fisher & Paykel Healthcare Adult Respiratory Ventilator Circuit-Dual Heated Catalogue Number RT210

This recall has been terminated (originally issued March 29, 2006).

Company
Fisher & Paykel Healthcare Inc
Recall Initiated
December 20, 2005
Posted
March 29, 2006
Terminated
April 19, 2012
Recall Number
Z-0696-06
Quantity
2175 boxes (21,750 pcs)
Firm Location
Laguna Hills, CA
Official Source
View on FDA website ↗

Reason for Recall

Risk of fire. The manufacturing process may have the potential to damage the insulation of the heater wire in circuits within the specified lot range. This may lead to a potential risk of fire.

Distribution

Nationwide in AR, CA, FL, IL, LA, MN, NC, OH, SD, TX, UT, & VA. International distribution

Lot / Code Info

Affected lots for all catalogue numbers: 041101-050331(inclusive).

Root Cause

Other

Action Taken

On December 20th, 2005 the firm sent the notification letter instructing consignees to check and destroy any stock in the affected lot range. A fax-back Device Recall Response Form was included with the letter and will be used for the company''s effectiveness checks.