Angio-Seal Vascular Closure Device. Ref 610109. 6F STS Platform. Sterilized by gamma radiation. For single use only. St. Jude Medical, Daig Division, Inc., Minnetonka, MN 55345-2126, USA.

This recall has been terminated (originally issued March 25, 2006).

Company
St. Jude Medical Daig Division
Recall Initiated
August 22, 2005
Posted
March 25, 2006
Terminated
July 22, 2006
Recall Number
Z-0677-06
Quantity
4160
Firm Location
Minnetonka, MN
Official Source
View on FDA website ↗

Reason for Recall

Mispackaging-Some Angio-Seal 6F STS Vascular Closure Devices from Lot no. 1146548 have been incorrectly packaged with a 0.035' guidewire instead of a 0.032' guidewire.

Distribution

International Only-Belgium, China, Guadalajara, New Zealand and Thailand.

Lot / Code Info

Lot number 1146548

Root Cause

Other

Action Taken

Letters were delivered to the hospitals and/or doctors in possession of the product instructing them to return it.