Angio-Seal Vascular Closure Device. Ref 610109. 6F STS Platform. Sterilized by gamma radiation. For single use only. St. Jude Medical, Daig Division, Inc., Minnetonka, MN 55345-2126, USA.

Company
St. Jude Medical Daig Division
Recall Initiated
August 22, 2005
Posted
March 25, 2006
Terminated
July 22, 2006
Recall Number
Z-0677-06
Quantity
4160
Firm Location
Minnetonka, MN

Reason for Recall

Mispackaging-Some Angio-Seal 6F STS Vascular Closure Devices from Lot no. 1146548 have been incorrectly packaged with a 0.035' guidewire instead of a 0.032' guidewire.

Distribution

International Only-Belgium, China, Guadalajara, New Zealand and Thailand.

Lot / Code Info

Lot number 1146548

Root Cause

Other

Action Taken

Letters were delivered to the hospitals and/or doctors in possession of the product instructing them to return it.