Angio-Seal Vascular Closure Device. Ref 610109. 6F STS Platform. Sterilized by gamma radiation. For single use only. St. Jude Medical, Daig Division, Inc., Minnetonka, MN 55345-2126, USA.
- Company
- St. Jude Medical Daig Division
- Recall Initiated
- August 22, 2005
- Posted
- March 25, 2006
- Terminated
- July 22, 2006
- Recall Number
- Z-0677-06
- Quantity
- 4160
- Firm Location
- Minnetonka, MN
Reason for Recall
Mispackaging-Some Angio-Seal 6F STS Vascular Closure Devices from Lot no. 1146548 have been incorrectly packaged with a 0.035' guidewire instead of a 0.032' guidewire.
Distribution
International Only-Belgium, China, Guadalajara, New Zealand and Thailand.
Lot / Code Info
Lot number 1146548
Root Cause
Other
Action Taken
Letters were delivered to the hospitals and/or doctors in possession of the product instructing them to return it.
More recalls by St. Jude Medical Daig Division
FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical,...
Mar 1, 2007
The affected device is labeled Angio-Seal Vascular Closure Device 6F VIP Platform, model 610130. Th...
Nov 7, 2006