FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and dilator; Product # 406850.

Company
St. Jude Medical Daig Division
Recall Initiated
October 5, 2006
Posted
March 1, 2007
Terminated
May 25, 2007
Recall Number
Z-0535-2007
Quantity
40 Units
Firm Location
Minnetonka, MN

Reason for Recall

Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.

Distribution

Nationwide.

Lot / Code Info

Lot number 1239904

Root Cause

Other

Action Taken

The affected customers were notified on 10/5/06 by letter.