FastCath Transseptal Catheter Introducers with Hemostasis Valve, DAIG division of St. Jude Medical, Minnetonka, MN, consists of a radiopaque sheath and dilator; Product # 406850.

This recall has been terminated (originally issued March 1, 2007).

Company
St. Jude Medical Daig Division
Recall Initiated
October 5, 2006
Posted
March 1, 2007
Terminated
May 25, 2007
Recall Number
Z-0535-2007
Quantity
40 Units
Firm Location
Minnetonka, MN
Official Source
View on FDA website ↗

Reason for Recall

Limited number of FastCath Transseptal Catheter Introducers (REF #406850) are susceptible to the irrigation extension tube on the side port becoming disconnected when a pull force is applied.

Distribution

Nationwide.

Lot / Code Info

Lot number 1239904

Root Cause

Other

Action Taken

The affected customers were notified on 10/5/06 by letter.