Artiste MV SA Linear Accelerator with syngo RT Therapist Version 4.1 build 67 and build 122. Manufactured by Siemens Medical Solutions Oncology Care Systems, Concord, CA.

This recall has been terminated (originally issued January 29, 2009).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
October 23, 2008
Posted
January 29, 2009
Terminated
December 9, 2010
Recall Number
Z-0927-2009
Quantity
36 units
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

If a plan contains a sequence of beams, including at least 1 ARC treatment beam and if during the setup stage of a beam following an ARC beam in the delivery sequence a termination occurs, and the user selects the immediate resumption feature, syngo RT Therapist Express systems will display and record a second delivery of a prior beam and not the delivery of the current beam in certain circumstanc

Distribution

Worldwide Distribution including USA and countries of Netherlands, Germany, Poland, Norway, Australia, Hungary, Belgium, New Zealand, France, Malaysia, Spain, and Canada.

Lot / Code Info

Part Number 8139789, Serial Numbers: 5190, 5350, 5355, 5364, 5367, 5374, 5377, 5398, 5405, 5410, 5419, 5424, 5165, 5166, 5202, 5238, 5298, 5306, 5342, 5352, 5365, 5366, 5373, 5378, 5379, 5380, 5397, 5399, 5400, 5401, 5402, 5409, 5414, 5416, 5423 and 5425.

Root Cause

Software design

Action Taken

An Urgent Medical Device Customer Safety Letter dated October 23, 2008 was issued to consignees via certified mail. The letter identified the issue and preventive measures for consignees. Monitoring will be done via return cards. Follow-up to be conducted via telephone.