SIEMENS brand ARTISTE syngo RT Therapist (RTT) version 4.3, ARTISTE Model, Part No. 08162815, Digital linear accelerators, MOSAIQ OIS system, Distributed by SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Manufactured by SIEMENS AG, MEDICAL SOLUTIONS, GERMANY.

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
January 17, 2013
Posted
March 5, 2013
Terminated
March 5, 2013
Recall Number
Z-0907-2013
Quantity
5 units
Firm Location
Concord, CA

Reason for Recall

Siemens became aware of a potential safety issue when using the "In-session resumption" feature of the syngo RT Therapist (RTT) 4.3 software. The RTT will send again the previously completed beam or segment for delivery to the control console, thus repeating the already delivered Monitor Units of the same beam or segment.

Distribution

5 devices, all located outside of the United States in Germany.

Lot / Code Info

All codes/units of ARTISTE model.

Root Cause

Software design

Action Taken

The firm issued "Urgent Medical Device Correction" letters via mail to all its direct consignees, informing them of the affected product issue and providing instructions on the recall. The firm stated that the release of this UI TH005/13/S to distribute the Customer Safety Advisory Notice began on January 17, 2013.