Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131

This recall has been terminated (originally issued May 11, 2005).

Company
Del Medical Systems Group
Recall Initiated
March 10, 2005
Posted
May 11, 2005
Terminated
November 14, 2005
Recall Number
Z-0816-05
Quantity
31 stands
Firm Location
Franklin Park, IL
Official Source
View on FDA website ↗

Reason for Recall

The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.

Distribution

Nationwide, Mexico and Thailand.

Lot / Code Info

All Phantom Model Tube Stands with squared corners on all four corners of the stand. Catalog #501031, serial numbers 00743-0204, 00778-0204, 00997-0304, 01001-0304, 01103-0304, 01293-0304, 01371-0304, 01425-0404, 01452-0404, 01470-0404, 01475-0404, 01575-0404, 01592-0404, 01700-0404, 01704-0404, 01790-0404, 01841-0404, 01851-0404, 01898-0504, 02019-0504, 02023-0504, 02051-0504, 02096-0504, 02178-0504, 02188-0504, 02310-0604, 02357-0604, 02429-0604, 02517-0604, 02534-0604, 02584-0604, 02671-0604, 02752-0704, 02760-0704, 02776-0704, 02839-0704, 02891-0704, 02928-0704, 02938-0704, 02945-0704, 02959-0704, 02980-0704, 02991-0704, 03076-0704, 03084-0704, 03122-0704, 03375-0804, 03455-0804, 03476-0804, 03518-0804, 03543-0804, 03570-0904, 03579-0904, 03636-0904, 03700-0904, 03708-0904, 03722-0904, 03740-0904, 03952-0904, 04037-0904, 04134-1004, 04183-1004, 04347-1004, 04380-1004, 04407-1004, 04477-1004, 04481-1004, 04491-1004, 04524-1004, 04534-1004, 04592-1104, 04600-1104, 04605-1104, 04648-1104, 04694-1104, 04699-1104, 04703-1104, 0439-1104, 04783-1104, 04862-1104, 05190-1204, 05194-1204, 05198-1204, 05216-1204, 05364-1204, 05414-1204, 05418-1204, 05439-1204, 00283-0105, 00340-0105, 00481-0105, 00571-0205, 00575-0205, 00617-0205, 00649-0205, 00763-0205, 00841-0205, 00858-0205, 00891-0205, 00900-0205, 00913-0205

Root Cause

Other

Action Taken

Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.