DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)

This recall has been terminated (originally issued October 19, 2006).

Company
Del Medical Systems Group
Recall Initiated
October 3, 2006
Posted
October 19, 2006
Terminated
January 24, 2008
Recall Number
Z-0049-2007
Quantity
53 units
Firm Location
Franklin Park, IL
Official Source
View on FDA website ↗

Reason for Recall

The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.

Distribution

Nationwide and Thailand

Lot / Code Info

Generator serial numbers: 03460-1003, 03477-1003, 03499-1003, 03550-1003, 03647-1003, 03704-1003, 03734-1003, 03744-1003, 03759-1003, 03765-1003, 03823-1003, 03938-1103, 03941-1103, 03953-1003, 04227-1103, 04282-1203, 04303-1203, 04330-1203, 04477-1203, 04497-1203, 04666-1203, 04704-1203, 00013-0104, 00035-0104, 00107-0104, 00122-0104, 00127-0104, 00188-0104, 00192-0104, 00332-0104, 00336-0104, 00405-0104, 00417-0104, 00430-0104, 00474-0104, 00545-0204, 00555-0204, 00568-0204, 00709-0204, 00719-0204, 00741-0204, 00754-0204, 00776-0204, 00995-0304, 00999-0304, 01101-0304, 01291-0304, 01295-0304, 01369-0304, 01423-0404, 01450-0404, 01468-0404, 01473-0404

Root Cause

Other

Action Taken

Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.