FloWire Doppler Guide wires. Volcano Corporation, Rancho Cardova, Ca. The Volcano FloWire Doppler guide wire is a single use guide wire intended for use in all blood vessels including both coronary and peripheral arteries to measure blood flow velocities during diagnostics angiography and/or interventional procedures.

Company
Volcano Corporation
Recall Initiated
October 6, 2009
Posted
February 1, 2012
Terminated
February 1, 2012
Recall Number
Z-0935-2012
Quantity
14 in US; 42 to EU and 15 to Japan
Firm Location
Rancho Cordova, CA

Reason for Recall

Reports of no signal displayed on the monitor when FloWire Doppler Guide wires were connected to the Combomap.

Distribution

Worldwide Distribution - USA including MI, AZ and FL and the countries of Japan and Europe

Lot / Code Info

Part number 1400; Serial numbers 011-0318-002; 011-01323-012; 011-01319-001; 011-01319-002, 011-01319-005, 011-01319-006; 011-01319-007; 011-01318-001, 011-01321-003, 011-01321-006; 011-01321-008; 011-01321-009; 011-01322-001, 011-01323-010.

Root Cause

Employee error

Action Taken

Volcano Corporation sent recall letters dated October 6, 2009, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and fax the attached Customer Reply Letter so that an RMA could be issued for the return of the unused affected product, and any replacement or credit could be arranged. Customers were instructed to return the Customer Reply letter even if they no longer have the affected product in their inventory. For questions regarding the use of the product. customers were instructed to contact their Sales Representative. For questions regarding this recall call 916-281-2790.