Philips Volcano CORE M2 Vascular System, Part # 400-0100.17, Rx Only. Used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature.

Company
Volcano Corporation
Recall Initiated
March 22, 2018
Posted
May 17, 2018
Terminated
April 28, 2021
Recall Number
Z-1891-2018
Quantity
8 devices
Firm Location
Rancho Cordova, CA

Reason for Recall

The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.

Distribution

Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.

Lot / Code Info

Serial #: 126620101597, 126620101599, 126620101600, 126620100718, 126620102313, 126620102315, 126620102307, and 126620093077.

Root Cause

Other

Action Taken

Customers were notified on 03/21/2018. U.S. customers were notified via hand-delivered Urgent Medical Device Recall letters; foreign customer letter was mailed. The letter identified the affected device and the reason for the recall. Instructions included to stop usage of the device and to call remote support at 1-800-228-4728, to schedule a visit to replace the affected device. Customers were instructed to complete and return the customer response form or to provide the form to Philips representatives. Customers were also instructed to provide copies of the recall notification to all personnel within their organization who handled this device.