CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.

Company
Biosense Webster, Inc.
Recall Initiated
February 9, 2012
Posted
March 20, 2012
Terminated
September 6, 2012
Recall Number
Z-1230-2012
Quantity
134 units total (32 units in US)
Firm Location
Irwindale, CA

Reason for Recall

Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.

Distribution

Worldwide Distribution -- US, Japan, China, Ireland, Germany, Italy, Slovenia, Indonesia, Poland, Spain, United Kingdom, Switzerland, Belgium, and Canada.

Lot / Code Info

Catalog # FG-5400-00  Serial numbers: 11996, 11998, 11999, 12002, 12026, 12031, 12040, 12046, 12047, 12051, 12052, 12057, 12061, 12062, 12063, 12065, 12066, 12067, 12069, 12070, 12075, 12077, 12078, 12086, 12087, 12088, 12089, 12094, 12096, 12098, 12099, and 12111.

Root Cause

Nonconforming Material/Component

Action Taken

Three Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customers are to contact their Biosense Webster representative. A Biosense Webster representative will contact customers to schedule an appointment for the replacement of the Backplane card. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. If customers have questions, they can contact their BWI representative.