WaveCrest Left Atrial Appendage (LAA), Occlusion System (32 mm), Catalog Numbers WCR1523 and WCR2523.

Company
Biosense Webster, Inc.
Recall Initiated
May 7, 2018
Terminated
November 14, 2020
Recall Number
Z-2878-2018
Quantity
25 units
Firm Location
Irwindale, CA

Reason for Recall

When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.

Distribution

No U.S. Distribution. International Distribution only.

Lot / Code Info

All lots.

Root Cause

Component design/selection

Action Taken

Coherex Medical notified customers via "URGENT FIELD SAFETY NOTICE Coherex WaveCrest Left Atrial Appendage (LAA) Occlusion System (32 mm) Catalog No: WCR1523, Lot Number: All" letter dated May 7, 2018. The letter provided an overview, and important information about the product description and codes, lots subject to removal, why customers were being contacted, What actions are required/instructions, available assistance/contact information, and Reply Form.