Medtronic FlexCath, model 3FC12, Steerable Sheath. The FlexCath Steerable Sheath is a percutaneous catheter introducer fitted with a valve to allow for introduction, withdrawal, and swapping of catheters and wires while preventing air ingress and minimizing blood loss. A side-port with stopcock is integrated to allow continuous drip infusion, injection through the center lumen, flushing, aspiration, blood sampling and pressure monitoring. Rx only. Sterilized using ethylene oxide. Manufacturer: Medtronic CryoCath LP, Kirkland, Quebec H9H 5H3, Canada. Product Usage: The FlexCath Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.

This recall has been terminated (originally issued September 12, 2011).

Company
Medtronic Inc. Cardiac Rhythm Disease Management
Recall Initiated
July 15, 2011
Posted
September 12, 2011
Terminated
June 7, 2012
Recall Number
Z-3186-2011
Quantity
6712
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Medtronic CryoCath would like to inform you of an observation related to a potential leak in the hemostatic valve of the Flex Cath 12 Steerable Sheath, Model 3FC12. There wave been reports of blood and saline solution leaking out of, and air ingress into the hemostatic valve. This includes reported instances of the introduction of air bubbles through the valve during aspiration of the side port

Distribution

Worldwide Distribution - USA (nationwide) including the states of: AZ, CA, CO, FL, GA, IA, IL, IN, KY, LA, MA, MI, MO, MN, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI. and the countries of: Australia, Austria, Belgium, Canada, Czech, Denmark, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, Netherlands, Norway, Poland, Qatar, South Africa, Spain, Switzerland, and United Kingdom.

Lot / Code Info

15440, 15441, 15438, 15439, 62503, 97317, 97313, 62504   14180, 53903, 14191, 14182, 93588, 93590, 93593   46732, 46733, 50665, 89982, 89988, 53902, 53900   06055, 16889, 16890, 16891, 20820, 26063, 26065, 26066, 91369   06044, 06045, 06046, 06047, 06049, 06048, 06050, 06052, 06051, 06053, 06054   05825, 05826, 05892, 05893, 05894, 05924, 05925, 05926, 05941, 05942, 05943   05650, 05651, 05652   05483, 05484, 05537, 05538   05371, 05372, 05398, 05399, 05482   05270, 05271, 05353, 05272, 05354, 05355, 05370   05127, 05128, 05171, 05172, 05225, 05226, 05267, 05268, 05269   04985, 04986, 05036, 05037, 05038   97318, 37267, 85114, 90801, 90782, 90778, 37266, 37264, 37229   85820, 09623, 85818, 85819, 85814, 09624, 09625, 09627, 09626, 00721, 00722, 96490   23512, 00724, 23511, 23697, 23698, 00723, 35329, 35328, 35332, 35331, 38075, 38074, 35334, 61553, 61555, 61554   96439, 61556, 61557, 94235, 94240, 94238, 94233, 26365, 26375, 26378   26382, 26385, 26384, 36902, 36905, 36909, 36911, 51097, 26374, 71318    71318, 51121, 36906, 51119, 51122, 68315, 26374, 68316, 68319, 77406, 77405, 77408   98733, 22990, 26379, 22989, 88479, 88481, 26512, 37353, 37355

Root Cause

Pending

Action Taken

Medtronic Sales Reps began on July 15, 21011 hand delivering an "IMPORTANT MEDICAL DEVICE iNFORMATION" letter dated July 2011 to consignees. The letter identified the product, the problem, and the actions needed to be taken. The letter stated the rate of occurrence, described the potential clinical implications, and reinforced the importance of following the information found in the instructions for use as consignees continue to use the device. The letter also stated that Medtronic CryoCath is working urgently on a number of fronts to mitigate this observation and will continue to communicate with consignees as appropriate. For further information, please contact Medtronic VP of Quality at (763) 526-0576.