Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.

This recall has been terminated (originally issued September 17, 2015).

Company
Medtronic Inc. Cardiac Rhythm Disease Management
Recall Initiated
September 1, 2015
Posted
September 17, 2015
Terminated
April 8, 2016
Recall Number
Z-2777-2015
Quantity
103 units
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

An issue with a USB memory component contained within a subset of CryoConsoles can result in extended procedure time.

Distribution

Worldwide Distribution - US (nationwide) and Internationally to Algeria, Australia, France, Germany, Greece, Iceland, Ireland, Israel, Italy, Lebanon, Malaysia, Martinique, Netherlands, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.

Lot / Code Info

serial numbers: 5K0005, 5K0006, 5K0007, 5K0009, 5K0010, 5K0011, 5K0012, 5K0013, 5K0014, 5K0015, 5K0016, 5K0017, 5K0018, 5K0019, 5K0020, 5K0021, 5K0022, 5K0023, 5K0024, 5K0025, 5K0026, 5K0027, 5K0028, 5K0029, 5K0030, 5K0031, 5K0032, 5K0033, 5K0034, 5K0035, 5K0036, 5K0037, 5K0039, 5K0040, 5K0041, 5K0042, 5K0043, 5K0044, 5K0045, 5K0046, 5K0047, 5K0048, 5K0053, 5K0054, 5K0056, 5K0057, 5K0058, 5K0059, 5K0062, 5K0063, 5K0064, 5K0065, 5K0066, 5K0067, 5K0068, 5K0069, 5K0070, 5K0071, 5K0072, 5K0073, N5923, N5924, N5925, N5929, N5930, N5931, N5932, N5933, N6106, N6119, N6210, N6720, N7763, N7764, N7765, N7766, N7767, N7768, N7769, N7770, N7771, N7772, N7773, N7774, N7775, N7776, N7777, N7778, N7779, N7780, N7781, N7782, N7783, N7784, N7785, N7786, N7787, N8029, N8030, N8031, N8033, N8034, N8035, N8038, N8039, N8040, N8041, N8042, N8043, N8044, N8045, N8046, N8047, N8048, N8049, N8050, N8051, N8052, N8053, N8054, N8055, N8056, N8057, N8058, N8059, N8060, N8061, N8062, N8063, N8064, N8065, N8067, N8068, N8069, N8070, N8072, N8073, N8074.

Root Cause

Component design/selection

Action Taken

Medtronic began hand delivering an Urgent Medical Device Correction letter on September 1, 2015, to all affected customers. The letter described the issue. Customers were advised to continue using the device, per the operator's manual, and if a system notice appears, they are to carefully read and follow the instructions in the screen of the console. Medtronic personnel will track contact with the customer via Customer Contact Record. Customers with qeustions were instructed to contact their Medtronic AF Solutions representative. For questions regarding this recall call 1-877-464-2796 Technical Services.