Medtronic CryoConsole, models 106A3, 106E2, and 106A2-K For use in performing cardiac ablation procedures.
This recall has been terminated (originally issued September 17, 2015).
- Recall Initiated
- September 1, 2015
- Posted
- September 17, 2015
- Terminated
- April 8, 2016
- Recall Number
- Z-2777-2015
- Quantity
- 103 units
- Firm Location
- Saint Paul, MN
- Official Source
- View on FDA website ↗
Reason for Recall
An issue with a USB memory component contained within a subset of CryoConsoles can result in extended procedure time.
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Algeria, Australia, France, Germany, Greece, Iceland, Ireland, Israel, Italy, Lebanon, Malaysia, Martinique, Netherlands, Poland, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Lot / Code Info
serial numbers: 5K0005, 5K0006, 5K0007, 5K0009, 5K0010, 5K0011, 5K0012, 5K0013, 5K0014, 5K0015, 5K0016, 5K0017, 5K0018, 5K0019, 5K0020, 5K0021, 5K0022, 5K0023, 5K0024, 5K0025, 5K0026, 5K0027, 5K0028, 5K0029, 5K0030, 5K0031, 5K0032, 5K0033, 5K0034, 5K0035, 5K0036, 5K0037, 5K0039, 5K0040, 5K0041, 5K0042, 5K0043, 5K0044, 5K0045, 5K0046, 5K0047, 5K0048, 5K0053, 5K0054, 5K0056, 5K0057, 5K0058, 5K0059, 5K0062, 5K0063, 5K0064, 5K0065, 5K0066, 5K0067, 5K0068, 5K0069, 5K0070, 5K0071, 5K0072, 5K0073, N5923, N5924, N5925, N5929, N5930, N5931, N5932, N5933, N6106, N6119, N6210, N6720, N7763, N7764, N7765, N7766, N7767, N7768, N7769, N7770, N7771, N7772, N7773, N7774, N7775, N7776, N7777, N7778, N7779, N7780, N7781, N7782, N7783, N7784, N7785, N7786, N7787, N8029, N8030, N8031, N8033, N8034, N8035, N8038, N8039, N8040, N8041, N8042, N8043, N8044, N8045, N8046, N8047, N8048, N8049, N8050, N8051, N8052, N8053, N8054, N8055, N8056, N8057, N8058, N8059, N8060, N8061, N8062, N8063, N8064, N8065, N8067, N8068, N8069, N8070, N8072, N8073, N8074.
Root Cause
Component design/selection
Action Taken
Medtronic began hand delivering an Urgent Medical Device Correction letter on September 1, 2015, to all affected customers. The letter described the issue. Customers were advised to continue using the device, per the operator's manual, and if a system notice appears, they are to carefully read and follow the instructions in the screen of the console. Medtronic personnel will track contact with the customer via Customer Contact Record. Customers with qeustions were instructed to contact their Medtronic AF Solutions representative. For questions regarding this recall call 1-877-464-2796 Technical Services.