Varian Medical Systems Floorstand. An accessory of the Optical Guidance Platform, Model number HZ5 for use with a charged particle accelerator (such as the Clinac 600C).

This recall has been terminated (originally issued January 10, 2011).

Company
Varian Medical Systems Oncology Systems
Recall Initiated
January 26, 2009
Posted
January 10, 2011
Terminated
February 24, 2012
Recall Number
Z-0888-2011
Quantity
3 units
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

Potential patient impact -- limit switches for Floorstand may be inadequate for some models and allow for potential impact.

Distribution

Product was distributed to 3 consignees in Florida and Wisconsin.

Lot / Code Info

Codes HZ5071, HZ59014 and HZ59017.

Root Cause

Device Design

Action Taken

Varian sent an urgent device correction notification letter 01/26/2009 to all affected users, with a description of the problem and user corrective action steps via certified mail. The letter will also be distributed to Varian Sales, Marketing and Service organizations.