Varian brand C-Series Ximatron Radiotherapy Treatment Stimulator, Model H72, Product is manufactured and distributed by Varian Oncology Systems, Palo Alto, CA The product is indicated for use in radiation therapy simulation, using a fluoroscopic and/or radiographic x-ray system for visualizing the volume to be exposed during radiation therapy.

This recall has been terminated (originally issued January 11, 2011).

Company
Varian Medical Systems Oncology Systems
Recall Initiated
January 19, 2010
Posted
January 11, 2011
Terminated
February 24, 2012
Recall Number
Z-0892-2011
Firm Location
Palo Alto, CA
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for failure of a lead screw drive system or image intensifier lead screw drive assembly, if normal maintenance inspections are not followed correctly.

Distribution

Worldwide Distribution including USA, Australia, New Zealand, Europe, Southeast Asia, China, Japan, Latin America, and Canada.

Lot / Code Info

Listed are the affected system Serial Numbers. The last four digits represent the serial number of the device.  H720003 ,H720119 ,H720209 ,H720270 ,H720347 ,H720416 ,H720491 ,H720557 ,H720618, H720006 ,H720122 ,H720210 ,H720271 ,H720348 ,H720417 ,H720495 ,H720558 ,H720620, H720009 ,H720123 ,H720211 ,H720273 ,H720350 ,H720419 ,H720496 ,H720559 ,H720622, H720013 ,H720124 ,H720213 ,H720274 ,H720351 ,H720422 ,H720498 ,H720560 ,H720624, H720015 ,H720125 ,H720215 ,H720276 ,H720352 ,H720424 ,H720499 ,H720561 ,H720626, H720017 ,H720129 ,H720217 ,H720279 ,H720353 ,H720425 ,H720500 ,H720562 ,H720627, H720021 ,H720132 ,H720218 ,H720280 ,H720354 ,H720427 ,H720501 ,H720563 ,H720628, H720022 ,H720133 ,H720219 ,H720281 ,H720355 ,H720429 ,H720502 ,H720565 ,H720629, H720025 ,H720134 ,H720221 ,H720282 ,H720357 ,H720432 ,H720503 ,H720566 ,H720632, H720035 ,H720137 ,H720222 ,H720283 ,H720358 ,H720435 ,H720505 ,H720567 ,H720637, H720037 ,H720138 ,H720225 ,H720284 ,H720362 ,H720437 ,H720506 ,H720568 ,H720638, H720040 ,H720140 ,H720229 ,H720288 ,H720363 ,H720438 ,H720508 ,H720571 ,H720639, H720041 ,H720143 ,H720230 ,H720289 ,H720367 ,H720439 ,H720509 ,H720572 ,H720641, H720048 ,H720144 ,H720231 ,H720290 ,H720373 ,H720440 ,H720511 ,H720573 ,H720642, H720051 ,H720145 ,H720232 ,H720292 ,H720374 ,H720442 ,H720513 ,H720575 ,H720643, H720052 ,H720148 ,H720233 ,H720293 ,H720375 ,H720443 ,H720514 ,H720577 ,H720644, H720056 ,H720152 ,H720234 ,H720294 ,H720376 ,H720445 ,H720515 ,H720578 ,H720647, H720057 ,H720154 ,H720235 ,H720295 ,H720379 ,H720447 ,H720516 ,H720579 ,H720648, H720059 ,H720157 ,H720236 ,H720296 ,H720383 ,H720449 ,H720518 ,H720584 ,H720650, H720060 ,H720159 ,H720237 ,H720297 ,H720384 ,H720452 ,H720522 ,H720585 ,H720651, H720064 ,H720164 ,H720239 ,H720298 ,H720386 ,H720457 ,H720523 ,H720586 ,H720653, H720069 ,H720165 ,H720241 ,H720299 ,H720388 ,H720458 ,H720525 ,H720588 ,H720654, H720070 ,H720167 ,H720242 ,H720300 ,H720391 ,H720461 ,H720526 ,H720589 ,H720659, H720071 ,H720168 ,H720245 ,H720303 ,H720392 ,H720465 ,H720528 ,H720590 ,H720660, H720072 ,H720169 ,H720247 ,H720307 ,H720394 ,H720466 ,H720530 ,H720591 ,H720661, H720074 ,H720177 ,H720248 ,H720309 ,H720395 ,H720470 ,H720531 ,H720592 ,H720663, H720075 ,H720180 ,H720251 ,H720316 ,H720396 ,H720471 ,H720533 ,H720594 ,H720664, H720082 ,H720181 ,H720254 ,H720320 ,H720397 ,H720472 ,H720534 ,H720595 ,H720666, H720083 ,H720182 ,H720255 ,H720327 ,H720399 ,H720477 ,H720536 ,H720596 ,H720667, H720088 ,H720183 ,H720256 ,H720328 ,H720402 ,H720478 ,H720537 ,H720597 ,H720668, H720089 ,H720185 ,H720257 ,H720329 ,H720403 ,H720479 ,H720539 ,H720598 ,H720669, H720092 ,H720186 ,H720258 ,H720330 ,H720404 ,H720481 ,H720546 ,H720599 ,H720670, H720098 ,H720187 ,H720261 ,H720336 ,H720405 ,H720482 ,H720547 ,H720600 ,H720671, H720101 ,H720188 ,H720262 ,H720337 ,H720406 ,H720484 ,H720548 ,H720601 ,H720672, H720106 ,H720195 ,H720263 ,H720338 ,H720407 ,H720485 ,H720550 ,H720604 ,H720673, H720107 ,H720196 ,H720264 ,H720339 ,H720409 ,H720486 ,H720553 ,H720605 ,H720674, H720108 ,H720199 ,H720265 ,H720344 ,H720411 ,H720487 ,H720554 ,H720608 ,H720675, H720112 ,H720200 ,H720266 ,H720345 ,H720414 ,H720488 ,H720555 ,H720615 ,H720676, H720113 ,H720204 ,H720269 ,H720346 ,H720415 ,H720489 ,H720556 ,H720616 ,H720678,   H720679 ,H720743 ,H720805 ,H720858 ,H720905 ,H720955 ,H721005, H720680 ,H720744 ,H720806 ,H720859 ,H720906 ,H720956 ,H721006, H720681 ,H720746 ,H720807 ,H720860 ,H720907 ,H720957 ,H721007, H720682 ,H720747 ,H720808 ,H720861 ,H720908 ,H720958 ,H721008, H720683 ,H720748 ,H720809 ,H720862 ,H720909 ,H720959 ,H721009, H720684 ,H720749 ,H720810 ,H720863 ,H720910 ,H720960 ,H721010, H720686 ,H720750 ,H720811 ,H720864 ,H720911 ,H720961 ,H721011, H720688 ,H720752 ,H720812 ,H720865 ,H720913 ,H720962 ,H721012, H720689 ,H720754 ,H720813 ,H720866 ,H720914 ,H720963 ,H721013, H720692 ,H720756 ,H720814 ,H720867 ,H720915 ,H720964 ,H721014, H720693 ,H720759 ,H720815 ,H720868 ,H720916 ,H720965 ,H721015, H720694 ,H720760 ,H720816 ,H720869 ,H720918 ,H720966 ,H721016, H720697 ,H720761 ,H720817 ,H720870 ,H720919 ,H720967 ,H721017, H720699 ,H720763 ,H720818 ,H720871 ,H720920 ,H720968 ,H721018, H720701 ,H720766 ,H720819 ,H720872 ,H720921 ,H720969 ,H721019, H720702 ,H720767 ,H720820 ,H720873 ,H720922 ,H720974 ,H721020, H720704 ,H720769 ,H720822 ,H720874 ,H720923 ,H720975 ,H721022, H720705 ,H720770 ,H720823 ,H720875 ,H720925 ,H720976 ,H721023, H720706 ,H720773 ,H720824 ,H720876 ,H720926 ,H720977 ,H721024, H720707 ,H720774 ,H720825 ,H720877 ,H720927 ,H720978 ,H721025, H720711 ,H720775 ,H720826 ,H720878 ,H720928 ,H720979 ,H721026, H720712 ,H720776 ,H720827 ,H720879 ,H720929 ,H720980 ,H721027, H720714 ,H720777 ,H720829 ,H720880 ,H720930 ,H720981 ,H721028, H720715 ,H720778 ,H720830 ,H720881 ,H720931 ,H720982 ,H721029, H720716 ,H720780 ,H720831 ,H720882 ,H720932 ,H720983 ,H721030, H720718 ,H720781 ,H720832 ,H720883 ,H720933 ,H720984 ,H721031, H720719 ,H720782 ,H720833 ,H720884 ,H720935 ,H720985 ,H721032, H720720 ,H720783 ,H720834 ,H720885 ,H720936 ,H720986 ,H721033, H720721 ,H720784 ,H720835 ,H720886 ,H720937 ,H720987 ,H721034, H720722 ,H720786 ,H720836 ,H720888 ,H720938 ,H720988 ,H721035, H720723 ,H720787 ,H720840 ,H720889 ,H720940 ,H720989 ,H721036, H720724 ,H720788 ,H720841 ,H720890 ,H720941 ,H720990 ,H721037, H720725 ,H720790 ,H720842 ,H720891 ,H720942 ,H720991 ,H721038, H720726 ,H720792 ,H720843 ,H720892 ,H720943 ,H720992 ,H721039, H720727 ,H720795 ,H720844 ,H720894 ,H720945 ,H720993 ,H721040, H720728 ,H720796 ,H720847 ,H720895 ,H720946 ,H720995 ,H721041, H720729 ,H720797 ,H720848 ,H720896 ,H720947 ,H720996 ,H721042, H720732 ,H720798 ,H720849 ,H720897 ,H720948 ,H720997 ,H721043, H720733 ,H720799 ,H720851 ,H720898 ,H720949 ,H720998 ,H721044, H720734 ,H720800 ,H720852 ,H720899 ,H720950 ,H720999 ,H721045, H720737 ,H720801 ,H720854 ,H720901 ,H720951 ,H721000 ,H721046, H720739 ,H720802 ,H720855 ,H720902 ,H720952 ,H721001 ,H721047, H720741 ,H720803 ,H720856 ,H720903 ,H720953 ,H721002 ,H721048, H720742 ,H720804 ,H720857 ,H720904 ,H720954 ,H721004

Root Cause

Other

Action Taken

Varian sent an URGENT MEDICAL DEVICE CORRECTION letter via certified mail receipt on January 19, 2010, to all direct consignees. The letter identified the product, the problem, and the action the consignee should take. Consignees were instructed to have maintenance personnel perform the recommended measurements as described in the Varian Customer Technical Bulletin. Ensure that all measurements made as part of the procedure are recorded in the 'wear table' that is part of the procedure. Previous wear measurements should be compared to new measurements and any large increase in wear should be investigated immediately. Pay special attention to the "Recommendations based on wear measurements" table. Ensure that new wear measurements comply with the operating parameters described in the table. For any questions consignees are to contact Varian Medical Systems Customer Support Services at one of the locations described in the CTB. For questions regarding this recall call (650) 424-5731.