MOSAIQ Oncology Information System Intended Use : Electronic Patient Records Management

This recall has been terminated (originally issued March 7, 2011).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
January 11, 2011
Posted
March 7, 2011
Terminated
April 19, 2012
Recall Number
Z-1572-2011
Quantity
10 units
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

The potential for injury to a patient should the treatment data be incorrectly uploaded to patient records.

Distribution

International Only Distribution -- Germany and Japan

Lot / Code Info

Part # 10568605

Root Cause

Software design (manufacturing process)

Action Taken

Sent an Urgent Medical Device Correction Safety Advisory Notice to customers on 1/11/11. Letters will be delivered by Siemens Customer Service Group or sent by certified mail. Letter identified the affected product and stated there is a potential risk due to a problem identified by the manufacturer. Customers are asked to refer to the attached Safety Notice from Elekta IMPAC Software. A Siemens Service Engineer will contact customers site to arrange for a patient matching criteria audit to determine if you have a problem. Customers should include the Safety Advisory Notice including the Safety Notice from Elekta IMPAC Software in their product related documentation until further notice. If customers have any questions, they should contact their responsible SIEMENS Service Engineer or Application Specialist.