Custom Packs/Mexico, Medtronic Cardio-Kit, Sterilized Using Ethylene Oxide, Nonpyrogenic, Rx Only. Model # DLP6N48R3 & DLP6N48R4. Medtronic Intersept Custom Tubing Pack with Trillium Biosurface, Sterilized Using Ethylene Oxide, , Model # TL4F71R1. These cannulae are intended for collection of venous blood from the right side of the heart via the superior and inferior vena cava during cardiopulmonary bypass surgery up to six hours or less.

This recall has been terminated (originally issued January 18, 2012).

Company
Medtronic Inc. Cardiac Rhythm Disease Management
Recall Initiated
December 14, 2011
Posted
January 18, 2012
Terminated
May 22, 2014
Recall Number
Z-0813-2012
Quantity
106
Firm Location
Saint Paul, MN
Official Source
View on FDA website ↗

Reason for Recall

Instructions for Use for the DLP Single Stage Venous Cannulae are in the process of being updated. The current precaution statement referencing ventricular assist procedures will be identified as a contraindication in the updated Instructions for Use. DO NOT USE THIS PRODUCT FOR EXTENDED TERMS SUCH AS VENTRICULAR ASSIST PROCEDURES. Medtronic has received two reports of patient deaths due to p

Distribution

Worldwide Distribution-USA (nationwide) including Puerto Rico and the Virgin Islands and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macau, Macedonia, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Viet Nam, and Yemen.

Lot / Code Info

Model # / Lot #    DLP6N48R3 DLP6N48R4 TL4F71R1 11241044 11465359 11799944  11755661  11713167  11713167  11669547  11616953  11568176  11509030  11478537  11432935  11414606  11282018  11252748  11239976  11209356  11130765  11110871  11101609  11051824  11015290  10916961  10860025  10816466  10693674  10656545  10588143  10502496  10454472  10426945  10310272  10255607   11669547  11616953  11568176  11509030  11478537  11432935  11414606  11282018  11252748  11239976  11209356  11130765  11110871  11101609  11051824  11015290  10916961  10860025  10816466  10693674  10656545  10588143  10502496  10454472  10426945  10310272  10255607

Root Cause

Labeling design

Action Taken

Medtronic sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2011 to all affected customers. The letter identifies the product, problem, and the action to be taken by the customers. Contract Medtronic CardioVascular LifeLine Technical Support at 1-877-526-7890 for questions related to this issue.