Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015. Product Code: 2M8161. Product Usage: Intended for use in a wide variety of patient environments for adult, pediatric and neonatal patients. Electronic infusion pumps indicated for continuous or intermittent delivery of solution through clinically acceptable routes of administration, such as intravenous (IV), intra-arterial (IA), subcutaneous, epidural or irrigation of fluid spaces applications.

Company
Baxter Healthcare Corp. Rt
Recall Initiated
July 14, 2010
Posted
January 13, 2012
Terminated
January 13, 2012
Recall Number
Z-0744-2012
Quantity
2 units
Firm Location
Round Lake, IL

Reason for Recall

Baxter's Service Center did not perform the ground impedance and leakage current test which verify if any power overflow occurs and that it is properly discharged through the device electrical system.

Distribution

Nationwide distribution: Puerto Rico.

Lot / Code Info

Product Code: 2M8161, Serial Numbers 15051625CC, 15020263CC

Root Cause

Employee error

Action Taken

Baxter sent service technicians to the consignees location on July 14, 2010 and October 6, 2010 and conducted the ground impedance and leakage current test on the affected pumps. The pumps were found with specifications.