Baxter Clearlink System Buretrol Solution Set, 115" (2.9 m) with 150 mL Clearlink Burette (Ball-Valve Drip Chamber); a sterile Rx IV fluid pathway device. Product Usage: The device is used for the administration of fluids from a container into the patient's vascular system through a vascular access device.

Company
Baxter Healthcare Corp. Rt
Recall Initiated
September 7, 2012
Posted
October 9, 2012
Terminated
December 29, 2017
Recall Number
Z-0014-2013
Quantity
2,265,453 units
Firm Location
Round Lake, IL

Reason for Recall

Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.

Distribution

Worldwide Distribution - US Nationwide and Internationally to Canada, Costa Rica, Australia, New Zealand, Hong Kong, and the United Kingdom.

Lot / Code Info

product code 2C8819, all lots

Root Cause

Device Design

Action Taken

Baxter Healthcare Corp. sent an Urgent Product Recall letter dated September 7, 2012, via first class mail to all affected customers, informing them of the problem with the ball-valve, and instructing them to discontinue use of the affected product and to contact Baxter for instructions to return the affected product.. The customers were also requested to complete the attached customer reply form, confirming their receipt of the letter, and fax it to Baxter at the number provided on the form. Any questions regarding the communication were directed to Medical Information Services at Baxter at 1-800-933-0303.