Pinnacle Destination Guiding Sheath, 6 French, 90 cm, Straight Tip, Tuohy-Borst Valve, Lot MD24, MD31, ME07, & ME14. 510(k) K012812 Guiding Sheath is used to facilitate the introduction of interventional and diagnostic devices into the human vasculature, including but not limited to the renal arteries for the Renal Guiding Sheath and the carotid arteries for the Carotid Guiding Sheath.

This recall has been terminated (originally issued June 6, 2011).

Company
Terumo Medical Corporation
Recall Initiated
June 24, 2010
Posted
June 6, 2011
Terminated
July 7, 2011
Recall Number
Z-2415-2011
Quantity
214 units
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

Molding defect in destination product that render the device unusable.

Distribution

Worldwide Distribution - USA including AL, AK, AZ, TX, LA, MS, MN, MO, CA, GA, FL, HI, KS, IA, ID, IL, TN, NM, OR, OK, WA, and WI and the countries of Belgium, Brazil, and Panama

Lot / Code Info

RSC01

Root Cause

Equipment maintenance

Action Taken

Terumo Medical Corporation sent an Urgent Product Bulletin Recall letter dated June 24, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo requested customers to immediately discontinue use, and quarantine any available stock of the listed product codes. In order to facilitate the exchange of these devices, a Terumo Territory Manager contacted all customers by July 9th to confirm receipt of the Terumo Destination Bulletin and scheduled an appointment to visit each customer in person by July 31st to personally manage the exchange. For questions or for further information, call (800) 283-7866 and ask to speak with the Product Recall Coordinator.