SOLOPATH Re-Collapsible Access System, Model Numbers SR-1925, SR-1935, SR-2025, SR-2035, SR-2225, SR-2235, SR-2425, SR-2435

This recall has been terminated (originally issued May 28, 2019).

Company
Terumo Medical Corporation
Recall Initiated
April 26, 2019
Posted
May 28, 2019
Terminated
September 22, 2020
Recall Number
Z-1375-2019
Quantity
1431
Firm Location
Elkton, MD
Official Source
View on FDA website ↗

Reason for Recall

There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting in a loss of the smooth transition from the surface of the tip to the outer surface of the expandable sheath.

Distribution

The products were distributed US nationwide. The products were distributed to the following foreign countries: Australia, Belgium, Brazil, Canada, Portugal, Hong Kong, Peru, Singapore.

Lot / Code Info

Lot Codes:VG26 VH21 VL20 VL29 VN29 WP05 VG02 VG07 VG13 VK29 VL18 WA25 WD02 WE25 WF09 WK01 WK09 WP05 XA09 VG06 VH27 VK04 XA09 VH27 VK07 VL28 WL14 XA30 VF31 VG16 VH21 WE25 WH23 WK01 VG09 VG14 VL14 VP18 WN07 XA09

Root Cause

Device Design

Action Taken

Urgent Medical Device Recall letters dated 4/26/19 were sent to customers.