2.0 mL Cartridge/Syringe Piston component of Insulin Pump Insulin infusion

This recall has been terminated (originally issued May 10, 2011).

Company
Animas Corporation
Recall Initiated
February 25, 2011
Posted
May 10, 2011
Terminated
April 3, 2014
Recall Number
Z-1678-2011
Quantity
384,180
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

Action is being undertaken to address the occurrence of adverse events associated with leaking cartridges. Firm became aware of this issue through customer complaints from pump users reporting the cartridges were leaking.

Distribution

Worldwide Distribution - USA (nationwide) and France.

Lot / Code Info

Lot numbers -- US: B201575, B201576, B201581, B201582, Exp. 7/31/2012; B201583, Exp. 8/31/2012; FRANCE: B201580, Exp. 7/31/2012.

Root Cause

Material/Component Contamination

Action Taken

Animas Corporation sent an Urgent Recall letter dated February 22, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Change their cartridge immediately using the instructions provided. Return all of the unused affected product using the enclosed box. Shipments must be returned using the United States Postal Service. For technical questions call Animas Customer Support at 1-855-254-5668 or call the toll free number at 1-8777-937-7867. For any questions about returning the product call 1-877-280-2339. Customers were instructed to notify the firm by logging on to the following website: www.animasnotification.com or call the Call Center at 1-877-280-2339 if they have no product to return.