Animas(R) 2020, Animas(R) IR 1250 (1200 plus), Animas(R) IR 1200 Insulin Infusion Pumps Indicated for continuous subcutaneous infusion of insulin.
This recall has been terminated (originally issued February 17, 2013).
- Company
- Animas Corporation
- Recall Initiated
- December 17, 2012
- Posted
- February 17, 2013
- Terminated
- December 24, 2013
- Recall Number
- Z-0836-2013
- Quantity
- 37,998
- Firm Location
- West Chester, PA
- Official Source
- View on FDA website ↗
Reason for Recall
Due to an internal calendar date limit (December 31, 2015) these pumps will cease to operate on January 1, 2016, and will result in a call service alarm that is displayed on the pump display.
Distribution
Worldwide Distribution-USA (nationwide) including Washington, DC, Puerto Rico, US Virgin Island and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Austria, Australia, Bermuda, Canada, New Zealand, Czech Republic, Italy, France, Finland, Hungary, Israel, United Kingdom, Spain, Sweden, Norway, Denmark, Germany, Mexico and Ireland.
Lot / Code Info
All Animas 2020, IR1250 and IR1200 Pump Models - Part Numbers 101370-00, 101371-00, 101372-00, 101374-00, 101375-00, 100380-00, 100381-00, 100382-00, 100384-00, 100385-00, 100450-00, 100210-00, 100211-00, 100212-00, 100250-00, 100251-00, 100252-00, 100200-00, 100201-00, 100202-00, 100170-00, 100171-00 and 100172-00.
Root Cause
Device Design
Action Taken
The firm initiated their recall of this product on December 17, 2012 by sending a letter to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Animas Customer Support Department at (866) 796-6373 for questions or concerns.