COBAS INTEGRA Creatinine Jaffe for use on the INTEGRA 400, 700 and 800 analyzers; system ID # 0764345, catalog # 20764345

This recall has been terminated (originally issued March 21, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
February 13, 2003
Posted
March 21, 2003
Terminated
March 20, 2003
Recall Number
Z-0667-03
Quantity
1905
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

After a cassette change, patient results and controls may be up to 40% too low or too high due to calibration factor variation.

Distribution

United States

Lot / Code Info

lot 639106

Root Cause

Other

Action Taken

An urgent product removal bulletin dated 2/14/03 was sent to each affected customer advising them to discontinue use of this lot, to discard any product on hand, and to contact Roche for replacement.