ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
- Company
- Icu Medical Inc
- Recall Initiated
- December 15, 2025
- Posted
- January 16, 2026
- Recall Number
- Z-1120-2026
- Firm Location
- Lake Forest, IL
Reason for Recall
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Lot / Code Info
400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0
Root Cause
Software design
Action Taken
ICU Medical issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/15/2025 via mail, email, telephone, or hand delivery. The notice explained the issue, potential risk and requested the following: "Actions for Users Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification and the sequence of events described. If you do experience this issue, the following mitigations can be performed: " To program the piggyback flush, the clinician must either a) clear infusion program data and power cycle the pump or b) obtain a replacement pump. " The clinician may also continue with the piggyback infusion without using the piggyback flush feature. To manually flush the downstream line, attach and program a flush syringe or bag on Line 2 to flush the fluid from the downstream line at the piggyback rate before restarting the primary line." For device correction inquiries, contact https://icumed.custhelp.com/app/market-action. For technical support, contact: 1-(800)-241-4002, option 3; tsc.support@icumed.com. For complaints, contact: 1-844-654-7780 (M-F, 8:00 am 5:00 pm CT) or ProductComplaintsPP@icumed.com.