LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Company
Icu Medical Inc
Recall Initiated
November 24, 2025
Posted
December 19, 2025
Recall Number
Z-0966-2026
Quantity
12 customers
Firm Location
Lake Forest, IL

Reason for Recall

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

Distribution

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

Lot / Code Info

LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227

Root Cause

Software Design Change

Action Taken

On November 24, 2025, the firm notified customers via Urgent Medical Device Correction letters. Consignees were asked to take the following actions: 1. Ensure all users or potential users of LifeShield DLM 2.2.1 are immediately made aware of this notification and proposed mitigations. 2. Until a software update is released, limit any newly created or edited Concentration Limits to one decimal place of precision for limits less than 100. ICU Medical is developing a software update to correct this issue and will notify customers of the timing of the software update.