Cardinal Health Clean Stool Transportation System-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Catalog Number: CHB900312

Company
Meridian Bioscience Inc
Recall Initiated
October 19, 2023
Posted
December 8, 2023
Recall Number
Z-0521-2024
Quantity
38 Cases
Firm Location
Cincinnati, OH

Reason for Recall

Due to a manufacturing defect, certain Para-Pak vials may be at risk for leakage.

Distribution

Nationwide Foreign: Italy

Lot / Code Info

UDI-DI: 50885380144859* *The UDI information for this product is maintained by Cardinal Health. Lot Numbers: 510020Q Expiration Date: 09/15/2030

Root Cause

Process control

Action Taken

Meridan issued Urgent Medical Device Correction letters via email on 10/19/23 to Distributors and End-Users. Letter states reason for recall, health risk and action to take: Identify and segregate any remaining inventory. REQUIRED ACTIONS FOR PARA-PAK CLEAN VIAL, CATALOG 900312, LOT 509940Q: " Identify and segregate any remaining inventory. " Discard remaining Inventory of Para-Pak Clean Vial, Catalog 900312, Lot 509940Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support at 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining inventory. REQUIRED ACTIONS FOR PARA-PAK Zn-PVA/10% FORMALIN, CATALOG 301012, LOT 509130Q: " Identify and segregate any remaining inventory. " For remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q, inspect each individual vial set for signs of leakage. o Leakage would be able to be visually observed through inspection of the Ziplock-style bag containing the two vials of preservatives. Signs of leakage include identification of preservative outside of the plastic vials or at the bottom of the Ziplock-style bag. DO NOT remove vials that exhibit leaking from the Ziplock-style bag. Handle with care to prevent direct contact of the preservative with the skin or eyes. o Record the quantity of 2 vial sets that demonstrated signs of leakage. " Discard the remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining invento

More recalls by Meridian Bioscience Inc

View all recalls by this company →