Injection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A

Company
Karl Storz Endoscopy
Recall Initiated
October 24, 2023
Posted
December 18, 2023
Recall Number
Z-0572-2024
Quantity
990
Firm Location
El Segundo, CA

Reason for Recall

The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.

Distribution

US: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI OUS: Mexica, Canada, Germany

Lot / Code Info

REF/UDI-DI: 27030N/4048551257180, 27182A/4048551081457 All lot numbers.

Root Cause

Device Design

Action Taken

On 10/24/23, recall notices were mailed to representatives for medical product safety, users, operators, distributors who were asked to do the following: 1) Discontinue use of all affected Injection Needles. 2) Pass on the Recall Notice to all users of the affected products and all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients, or provide the recalling firm with a list of customers who received/may have received the products listed. 4) Contact Customer Support to arrange for the return of affected devices by calling 800-421-0837. 5) Return the completed Customer Reply Form via email to karlstorz5190@sedgwick.com For all other questions regarding this communication, please send an email to KSEAProductLabeling@karlstorz.com

More recalls by Karl Storz Endoscopy

View all recalls by this company →