KARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6, 34 cm, NON STERILE, RxONLY, CE 0123

Company
Karl Storz Endoscopy
Recall Initiated
December 19, 2024
Posted
January 30, 2025
Recall Number
Z-1062-2025
Quantity
185 units
Firm Location
El Segundo, CA

Reason for Recall

Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.

Distribution

US Nationwide distribution.

Lot / Code Info

All Lots/UDI:04048551231227

Root Cause

Labeling Change Control

Action Taken

On December 19, 2024, Karl Storz, issued an "Urgent: Medical Device Correction" notification to affected consignees via UPS. In addition to informing consignees about the recall, Karl Storz asked consignees to take the following actions: 1. Immediately discard any previous revisions of the IFU. 2. Access the updated e-IFU at the following link: https://www.karlstorz.com/de/en/eifu.htm?q=&country=US 3. Pass on this Urgent Medical Device Correction to all users of the recalled products and all other persons who need to be aware within your organization. 4. If you have or may have distributed the products listed, please identify and promptly notify those recipients with a copy of this notice, response form and updated e-IFU. 5. Return the completed Customer Reply Form by Fax or E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 6. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

More recalls by Karl Storz Endoscopy

View all recalls by this company →