VERO BIOTECH GENOSYL DS (Delivery System) Generation 3 Cassette, Model Number 602722-01; generates and delivers nitric oxide (NO) for inhalation at the point of use

This recall is currently active, issued December 15, 2023. It was issued by Vero Biotech, Llc.

Company
Vero Biotech, Llc
Recall Initiated
November 17, 2023
Posted
December 15, 2023
Recall Number
Z-0564-2024
Quantity
2 cassettes
Firm Location
Atlanta, GA
Official Source
View on FDA website ↗

Reason for Recall

Cassette did not meet an 1n-process flow criterion.

Distribution

US: AZ and FL

Lot / Code Info

Lot Code: W-0001-2023, Cassette serial numbers: GA01T707, GA01T108

Root Cause

Process control

Action Taken

VERO BIOTECH issued an URGENT: MEDICAL DEVICE REMOVAL notice to its consignees on 11/17/2023 via email, The notice explained the issue and advised the consignee of the removal of the device from their possession on 11/03/2023.