GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860
This recall has been terminated (originally issued June 23, 2021).
- Company
- Vero Biotech, Llc
- Recall Initiated
- May 19, 2021
- Posted
- June 23, 2021
- Terminated
- October 12, 2023
- Recall Number
- Z-1863-2021
- Quantity
- 465 units
- Firm Location
- Atlanta, GA
- Official Source
- View on FDA website ↗
Reason for Recall
Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.
Distribution
Distributed nationwide to TX, MD, VA, GA, FL, KS, LA, MN, and NC
Lot / Code Info
Model #: 601588-01; Serial #'s: G0803, G0807, G0808, G0814, G0149, G0157, G0183, G0191, G0195, G0198, G0202, G0560, G0600, G0693, G0729, G0734, G0772, G0775, G0138, G0147, G0155, G0159, G0172, G0185, G0203, G0209, G0282, G0344, G0385, G0398, G0408, G0415, G0569, G0601, G0756, G0323, G0358, G0379, G0383, G0386, G0405, G0334, G0399, G0413, G0463, G0158, G0174, G0285, G0305, G0466, G0719, G0135, G0194, G0205, G0281, G0362, G0407, G0573, G0574, G0578, G0579, G0582, G0583, G0587, G0590, G0591, G0594, G0596, G0597, G0598, G0603, G0609, G0611, G0680, G0708, G0110, G0111, G0131, G0137, G0140, G0176, G0207, G0212, G0232, G0237, G0246, G0251, G0253, G0262, G0263, G0266, G0280, G0449, G0453, G0608, G0642, G0644, G0663, G0688, G0712, G0717, G0739, G0745, G0747, G0762, G0764, G0765, G0793, G0125, G0129, G0133, G0169, G0181, G0257, G0292, G0645, G0804, G0180, G0293, G0311, G0322, G0338, G0341, G0342, G0353, G0480, G0499, G0506, G0519, G0521, G0522, G0565, G0681, G0735, G0738, G0748, G0754, G0760, G0770, G0773, G0773, G0782, G0223, G0227, G0235, G0236, G0347, G0352, G0355, G0359, G0378, G0382, G0175, G0273, G0476, G0562, G0792, G0162, G0163, G0189, G0607, G0143, G0160, G0164, G0214, G0215, G0283, G0284, G0296, G0298, G0303, G0313, G0315, G0319, G0329, G0333, G0335, G0370, G0375, G0495, G0509, G0510, G0511, G0550, G0571, G0588, G0589, G0593, G0646, G0672, G0689, G0723, G0724, G0725, G0736, G0744, G0771, G0778, G0795, G0204, G0218, G0269, G0272, G0343, G0580, G0193, G0294, G0300, G0302, G0308, G0465, G0518, G0567, G0190, G0278, G0279, g0286, G0297, G0374, G0393, G0728, G0307, G0309, G0312, G0318, G0320, G0406, G0434, G0643, G0229, G0230, G0231, G0240, G0460, G0469, G0488, G0489, G0493, G0496, G0500, G0501, G0599, G0605, G0610, G0647, G0648, G0650, G0651, G0653, G0654, G0655, G0656, G0658, G0661, G0667, G0668, G0670, G0671, G0677, G0678, G0686, G0691, G0694, G0697, G0702, G0707, G0713, G0715, G0716, G0718, G0726, G0730, G0733, G0741, G0755, G0806, G0813, G0818, G0261, G0270, G0274, G0275, G0094, G0363, G0369, G0430, G0435, G0440, G0444, G0452, G0475, G0487, G0490, G0508, G0537, G0549, G0234, G0380, G0389, G0402, G0409, G0483, G0497, G0576, G0584, G0585, G0277, G0346, G0390, G0423, G0443, G0445, G0447, G0451, G0454, G0455, G0457, G0458, G0459, G0461, G0464, G0467, G0474, G0479, G0481, G0486, G0492, G0504, G0512, G0515, G0516, G0517, G0520, G0524, G0527, G0528, G0529, G0531, G0532, G0533, G0536, G0539, G0542, G0545, G0546, G0547, G0551, G0552, G0553, G0554, G0557, G0558, G0559, G0561, G0570, G0581, G0602, G0710, G0743, G0749, G0752, G0768, G0787, G0289, G0371, G0357, G0372, G0391, G0701, G0186, G0327, G0337, G0568, G0604, G0675, G0679, G0682, G0692, G0700, G0703, G0396, G0397, G0228, G0245, G0288, G0387, G0414, G0418, G0420, G0437, G0439, G0535, G0572, G0211, G0268, G0365, G0544, G0592, G0252, G0349, G0606, G0192, G0417, G0485, G0507, G0523, G0563, G0146, G0166, G0304, G0306, G0314, G0332, G0339, G0403, G0422, G0424, G0428, G0429, G0431, G0432, G0436, G0450, G0473, G0491, G0426, G0427, G0301, G0425, G0706, G0720, G0722, G0727, G0789, G0797, G0799, G0805, G0665, G0684, G0696, G0714, G0740, G0796, G0367, G0595, G0721, G0731, G0732, G0109, G0124, G0136, G0139, G0331, G0468, G0566, G0810, G0816, G0381, G0767, G0779, G0790, G0737, G0763, G0766, G0774
Root Cause
Software design
Action Taken
The firm initiated their recall by visit on 04/07/2021. VERO Clinical Educators are going (or have gone) onsite to the hospitals to physically upgrade the software to v2.2.4. The correction of upgrading the software from v2.2.3 to v2.2.4 began on Wednesday April 7, 2021. The software upgrade is expected to be completed by Friday April 23, 2021. For questions contact VERO Biotech at 1-877-337-4118 or visit the VERO Biotech website at https://www.vero-biotech.com/.