RecallDepth

Vitality Polyaxial Screw, 7.5mmx 50mm, REF: 07.02000.118, used with Vitality and Vital Spinal Fixation System

This recall is currently active, issued January 4, 2024. It was issued by Zimmer Biomet Spine Inc..

Company
Zimmer Biomet Spine Inc.
Recall Initiated
October 16, 2023
Posted
January 4, 2024
Recall Number
Z-0639-2024
Quantity
3
Firm Location
Westminster, CO
Official Source
View on FDA website ↗

Reason for Recall

Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.

Distribution

US Nationwide distribution in the states of MN, NE.

Lot / Code Info

UDI-DI: 00889024002159, Lot: W834061

Root Cause

Nonconforming Material/Component

Action Taken

On 10/16/23, recall notices were sent to customers asking them to do the following: 1) Locate, quarantine, and return affected screws to the recalling firm. 2) Complete and return the inventory return certificate form via email to Alex.Hayden@ZimVie.com Recall-related questions or concerns can be directed to please email Alex.Hayden@ZimVie.com.

More recalls by Zimmer Biomet Spine Inc.

View all recalls by this company →