RecallDepth

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

This recall has been terminated (originally issued May 15, 2017).

Company
Zimmer Biomet Spine Inc.
Recall Initiated
May 10, 2017
Posted
May 15, 2017
Terminated
May 25, 2018
Recall Number
Z-2318-2017
Quantity
497 units
Firm Location
Westminster, CO
Official Source
View on FDA website ↗

Reason for Recall

Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle 90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

Distribution

Nationwide, military/VA/govt and foreign consignee (No Canadian consignees)

Lot / Code Info

Model No: 07.02066.001 (All lots); 07.02053.001 (All lots distributed prior to April 26, 2017), and 730M0017 (Lot 309622)

Root Cause

Device Design

Action Taken

Consignees will be notified on 5/15/17 via certified mail to return products.

More recalls by Zimmer Biomet Spine Inc.

View all recalls by this company →