Vasoview Hemopro Endoscopic Vessel Harvesting System, model VH-3000-W .
- Company
- Maquet Cardiovascular, Llc
- Recall Initiated
- December 1, 2023
- Posted
- January 12, 2024
- Recall Number
- Z-0723-2024
- Quantity
- 554 units (OUS)
- Firm Location
- Wayne, NJ
Reason for Recall
Risk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
Distribution
US: OUS: China, Brazil, Hong Kong.
Lot / Code Info
UDI-DI: 00607567700413, Lots: 3000283735, 3000293031, 3000310304, 3000333956.
Root Cause
Component design/selection
Action Taken
An Urgent Medical Device Correction notice dated December 1, 2023 was sent via FedEx 2-Day delivery. Please examine your inventory immediately to determine if you have any of the product codes and lot numbers listed in the notice and remove these from use. Return any unused/unexpired affected product to Maquet/Getinge. Please forward this information to all current and potential VasoView HemoPro EVH System users within your hospital / facility. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. If you have any questions, please contact your Maquet/Getinge representative or call the Maquet/Getinge Customer Support at 888-880-2874, Monday through Friday, between the hours of 6 AM to 5 PM (Pacific Standard Time). Whether or not your facility has affected product listed in the notice, please complete and sign the MEDICAL DEVICE REMOVAL - RESPONSE FORM to acknowledge that you have received the notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to VVHemoProParticulate2023.act@getinge.com or by faxing the form to 1-800-878-9105.