Heartstring III Proximal Seal System, 3.8 MM. Intravascular anastomosis occluder.

Company
Maquet Cardiovascular, Llc
Recall Initiated
August 15, 2025
Posted
September 18, 2025
Recall Number
Z-2587-2025
Quantity
28,331 units (16,615 US, 11,716 OUS)
Firm Location
Wayne, NJ

Reason for Recall

Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.

Distribution

Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.

Lot / Code Info

Model No. HSK-3038; UDI: 00607567700314¿; Serial No. 3000377764, 3000378278, 3000379222, 3000379987, 3000383569, 3000388598, 3000390377, 3000394076, 3000396875, 3000398149, 3000400694, 3000402932, 3000403566, 3000404026, 3000404996, 3000405699, 3000407693, 3000409172, 3000409619, 3000410091, 3000410583, 3000412000, 3000412963, 3000415599, 3000416141, 3000417121, 3000417711, 3000418572, 3000418772, 3000419866, 3000420311, 3000422320, 3000422757, 3000423694, 3000424888, 3000425725, 3000428927, 3000429846, 3000431259, 3000431934, 3000434091, 3000434862, 3000435356, 3000436032, 3000436446, 3000440202, 3000441040, 3000441162, 3000442134, 3000443538, 3000443910, 3000444136, 3000446374, 3000447343, 3000448483, 3000449214, 3000449897, 3000450561, 3000451914, 3000452532, 3000453432, 3000453804, 3000455728, 3000456346, 3000457969, 3000458434, 3000459087, 3000459889, 3000460034, 3000461587, 3000464463, 3000465229, 3000465676, 3000465679, 3000466111, 3000466878, 3000467559, 3000470706, 3000471492, 3000473348, 3000473350, 3000475032, 3000475033, 3000476158, 3000476611, 3000477574, 3000478410, 3000479294, 3000479465, 3000481661, 3000484284, 3000485833, 3000488966, 3000377764, 3000378278, 3000379222, 3000379987, 3000388598, 3000391049, 3000394076, 3000396875, 3000398149, 3000400694, 3000403566, 3000404026, 3000404996, 3000405699, 3000409172, 3000409619, 3000410091, 3000412000, 3000412963, 3000415599, 3000417121, 3000417752, 3000418572, 3000419866, 3000420311, 3000422320, 3000423694, 3000424888, 3000425725, 3000428927, 3000431259, 3000431934, 3000434091, 3000434862, 3000435356, 3000436032, 3000436109, 3000436446, 3000437664, 3000440202, 3000440757, 3000441162, 3000442134, 3000442635, 3000443538, 3000443910, 3000444136, 3000446374, 3000447140, 3000447343, 3000448483, 3000449214, 3000449897, 3000450292, 3000451914, 3000452532, 3000452651, 3000453432, 3000453804, 3000455728, 3000456197, 3000456346, 3000457780, 3000457969, 3000458434, 3000459889, 3000460034, 3000461587, 3000464463, 3000465228, 3000465229, 3000465676, 3000465679, 3000466111, 3000466878, 3000467559, 3000470706, 3000471492, 3000473348, 3000475032, 3000475033, 3000476158, 3000477574, 3000478410, 3000479294, 3000479465, 3000481661, 3000484284, 3000484896, 3000488349, 3000488742, 3000488966.

Root Cause

Under Investigation by firm

Action Taken

Consignees were sent an URGENT MEDICAL DEVICE CORRECTION notification dated August 2025. The notification provides instructions for consignees to carry out in the instance of failure of the Heartstring Seal to load, failure of the Heartsring III Seal to deploy into the aortotomy, and failure of the deployed Heartstring III Seal to provide adequate hemostasis. Consignees are asked to forward the recall notification to individuals within their organization who use the recalled devices and to organizations to which product was further distributed. Consignees are to return the completed response form to recallresponses.qrc@getinge.com. Questions about this recall can be directed to Getinge Customer Service at 1-888-880-2874.