Smiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.

This recall is currently active, issued October 31, 2023. It was issued by Smiths Medical Asd Inc..

Company
Smiths Medical Asd Inc.
Recall Initiated
August 22, 2023
Posted
October 31, 2023
Recall Number
Z-0205-2024
Firm Location
Minneapolis, MN
Official Source
View on FDA website ↗

Reason for Recall

The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026).

Distribution

Distribution was made to AZ, CA, FL, HI, IL, IN, MA, MN, MO, MS, NC, NJ, OH, PA, SC, TX, VA, and WA. There was no government/military distribution. Foreign distribution was made to Canada.

Lot / Code Info

Lot #4125816, exp. 3/17/2026, UDI-DI 15021312005769.

Root Cause

Labeling mix-ups

Action Taken

On 8/22/2023, a third-party recall notification service issued recall notifications to the direct accounts on behalf of the recalling firm via UPS ground. The letter explained the reason for the notification and embedded pictures of the incorrect and correct labels into the letter. The letter also provided the potential risk and required actions for users. The potential risk involves the potential for a customer to assess the product as expired, resulting in the potential for a delay in the availability to use the tracheostomy device. If the customer did not identify the incorrect device unit carton/secondary packaging labeling upon receipt and provided the items to clinicians, the individual device tray/primary packaging would still allow correct identification of the expiration date at the point of use, as the expiry date on the individual unit is correctly labeled. The required actions included (1) informing potential users within their organization of the notification and to complete and return the attached response form via email within ten days of receipt to acknowledge if they have the affected product and/or if it has already been used; and (2) if the account has further distributed the product, immediately notify those customers of the field correction. The letter instructs that if the product has not already been used, the product must be used before the correct expiry of March 17, 2026.